2.
Measurement Process Characterization
2.3. Calibration 2.3.4. Catalog of calibration designs 2.3.4.2. Drift-elimination designs for gage blocks
|
|||
Doiron 3-9 Design
OBSERVATIONS 1 1 1
Y(1) + -
Y(2) - +
Y(3) + -
Y(4) - +
Y(5) - +
Y(6) + -
Y(7) - +
Y(8) - +
Y(9) + -
RESTRAINT +
CHECK STANDARD +
DEGREES OF FREEDOM = 7
SOLUTION MATRIX
DIVISOR = 9
OBSERVATIONS 1 1 1
Y(1) 0 -2 -1
Y(2) 0 -1 1
Y(3) 0 -1 -2
Y(4) 0 2 1
Y(5) 0 1 2
Y(6) 0 1 -1
Y(7) 0 2 1
Y(8) 0 -1 1
Y(9) 0 -1 -2
R(1) 9 9 9
FACTORS FOR COMPUTING REPEATABILITY STANDARD DEVIATIONS
NOM FACTOR
1 1 1
1 0.0000 +
1 0.4714 +
1 0.4714 +
1 0.4714 +
Explanation of notation and
interpretation of tables
| |||