2.
Measurement Process Characterization
2.3. Calibration 2.3.4. Catalog of calibration designs 2.3.4.2. Drift-elimination designs for gage blocks
|
|||
Doiron 4-8 Design
OBSERVATIONS 1 1 1 1
Y(1) + -
Y(2) + -
Y(3) - +
Y(4) + -
Y(5) - +
Y(6) - +
Y(7) + -
Y(8) - +
RESTRAINT +
CHECK STANDARD +
DEGREES OF FREEDOM = 5
SOLUTION MATRIX
DIVISOR = 8
OBSERVATIONS 1 1 1 1
Y(1) 0 -3 -2 -1
Y(2) 0 1 2 -1
Y(3) 0 1 2 3
Y(4) 0 1 -2 -1
Y(5) 0 3 2 1
Y(6) 0 -1 -2 1
Y(7) 0 -1 -2 -3
Y(8) 0 -1 2 1
R* 8 8 8 8
R* = Value of the reference standard
FACTORS FOR REPEATABILITY STANDARD DEVIATIONS
NOM FACTOR
1 1 1 1
1 0.0000 +
1 0.6124 +
1 0.7071 +
1 0.6124 +
1 0.6124 +
Explanation of notation and
interpretation of tables
| |||