2.
Measurement Process Characterization
2.3. Calibration 2.3.4. Catalog of calibration designs 2.3.4.2. Drift-elimination designs for gage blocks
|
|||
Doiron 7-14 Design OBSERVATIONS 1 1 1 1 1 1 1 Y(1) + - Y(2) - + Y(3) + - Y(4) + - Y(5) + - Y(6) - + Y(7) + - Y(8) + - Y(9) + - Y(10) - + Y(11) - + Y(12) - + Y(13) - + Y(14) - + RESTRAINT + CHECK STANDARD + DEGREES OF FREEDOM = 8 PARAMETER VALUES DIVISOR = 1015 OBSERVATIONS 1 1 1 1 1 1 1 Y(1) 0 -406 -203 -203 -203 -203 -203 Y(2) 0 0 -35 -210 35 210 0 Y(3) 0 0 175 35 -175 -35 0 Y(4) 0 203 -116 29 -116 29 -261 Y(5) 0 -203 -229 -214 -264 -424 -174 Y(6) 0 0 -175 -35 175 35 0 Y(7) 0 203 -61 -221 -26 -11 29 Y(8) 0 0 305 90 130 55 -145 Y(9) 0 0 220 15 360 -160 145 Y(10) 0 203 319 174 319 174 464 Y(11) 0 -203 26 11 61 221 -29 Y(12) 0 0 -360 160 -220 -15 -145 Y(13) 0 203 264 424 229 214 174 Y(14) 0 0 -130 -55 -305 -90 145 R* 1015 1015 1015 1015 1015 1015 1015 R* = Value of the reference standard FACTORS FOR REPEATABILITY STANDARD DEVIATIONS NOM FACTOR 1 1 1 1 1 1 1 1 0.0000 + 1 0.6325 + 1 0.7841 + 1 0.6463 + 1 0.7841 + 1 0.6463 + 1 0.6761 + 1 0.6325 +Explanation of notation and interpretation of tables |