2.
Measurement Process Characterization
2.3. Calibration 2.3.4. Catalog of calibration designs 2.3.4.3. Designs for electrical quantities
|
|||
Design for 4 references zeners and 2 test zeners.
ZENERS
OBSERVATIONS 1 1 1 1 1 1
Y(1) + -
Y(2) + -
Y(3) + -
Y(4) + -
Y(5) + -
Y(6) + -
Y(7) + -
Y(8) + -
Y(9) - +
Y(10) - +
Y(11) - +
Y(12) - +
Y(13) - +
Y(14) - +
Y(15) - +
Y(16) - +
RESTRAINT + + + +
CHECK STANDARD + -
DEGREES OF FREEDOM = 10
SOLUTION MATRIX
DIVISOR = 16
OBSERVATIONS 1 1 1 1 1 1 P
Y(1) 3 -1 -1 -1 -2 0 1
Y(2) 3 -1 -1 -1 0 -2 1
Y(3) -1 3 -1 -1 -2 0 1
Y(4) -1 3 -1 -1 0 -2 1
Y(5) -1 -1 3 -1 -2 0 1
Y(6) -1 -1 3 -1 0 -2 1
Y(7) -1 -1 -1 3 -2 0 1
Y(8) -1 -1 -1 3 0 -2 1
Y(9) 1 1 1 -3 2 0 1
Y(10) 1 1 1 -3 0 2 1
Y(11) 1 1 -3 1 2 0 1
Y(12) 1 1 -3 1 0 2 1
Y(13) 1 -3 1 1 2 0 1
Y(14) 1 -3 1 1 0 2 1
Y(15) -3 1 1 1 2 0 1
Y(16) -3 1 1 1 0 2 1
R* 4 4 4 4 4 4 0
R* = AVERAGE VALUE OF 4 REFERENCE STANDARDS
P = LEFT-RIGHT EFFECT
FACTORS FOR COMPUTING STANDARD DEVIATIONS
V FACTORS ZENERS
1 1 1 1 1 1 P
1 0.4330 +
1 0.4330 +
1 0.4330 +
1 0.4330 +
1 0.3536 +
1 0.3536 +
1 0.2500 +
| |||
| Explanation of notation and interpretation of tables | |||