2.
Measurement Process Characterization
2.3. Calibration 2.3.4. Catalog of calibration designs 2.3.4.3. Designs for electrical quantities
|
|||
Design for 4 references and 3 test zeners.
ZENERS
OBSERVATIONS 1 1 1 1 1 1 1
Y(1) - +
Y(2) - +
Y(3) + -
Y(4) + -
Y(5) + -
Y(6) + -
Y(7) - +
Y(8) - +
Y(9) - +
Y(10) - +
Y(11) - +
Y(12) - +
Y(13) + -
Y(14) + -
Y(15) + -
Y(16) + -
Y(17) + -
Y(18) - +
RESTRAINT + + + +
CHECK STANDARD + -
DEGREES OF FREEDOM = 11
SOLUTION MATRIX
DIVISOR = 1260
OBSERVATIONS 1 1 1 1 1 1 1 P
Y(1) -196 196 -56 56 0 0 0 70
Y(2) -160 -20 160 20 0 0 0 70
Y(3) 20 160 -20 -160 0 0 0 70
Y(4) 143 -53 -17 -73 0 0 -315 70
Y(5) 143 -53 -17 -73 0 -315 0 70
Y(6) 143 -53 -17 -73 -315 0 0 70
Y(7) 53 -143 73 17 315 0 0 70
Y(8) 53 -143 73 17 0 315 0 70
Y(9) 53 -143 73 17 0 0 315 70
Y(10) 17 73 -143 53 0 0 315 70
Y(11) 17 73 -143 53 0 315 0 70
Y(12) 17 73 -143 53 315 0 0 70
Y(13) -73 -17 -53 143 -315 0 0 70
Y(14) -73 -17 -53 143 0 -315 0 70
Y(15) -73 -17 -53 143 0 0 -315 70
Y(16) 56 -56 196 -196 0 0 0 70
Y(17) 20 160 -20 -160 0 0 0 70
Y(18) -160 -20 160 20 0 0 0 70
R* 315 315 315 315 315 315 315 0
R* = Average value of the 4 reference zeners
P = left-right effect
FACTORS FOR REPEATABILITY STANDARD DEVIATIONS
V K1 1 1 1 1 1 1 1
1 0.5000 +
1 0.5000 +
1 0.5000 +
2 0.7071 + +
3 0.8660 + + +
0 0.5578 + -
| |||
| Explanation of notation and interpretation of tables | |||