2.
Measurement Process Characterization
2.3. Calibration 2.3.6. Instrument calibration over a regime
|
|||
Purpose | The purpose of creating the calibration curve is to correct future measurements made with the same instrument to the correct units of measurement. The calibration curve can be applied many, many times before it is discarded or reworked as long as the instrument remains in statistical control. Chemical measurements are an exception where frequently the calibration curve is used only for a single batch of measurements, and a new calibration curve is created for the next batch. | ||
Notation |
The notation for this section is as follows:
|
||
Procedure |
To apply a correction to a future measurement, |
||
Linear calibration line |
The inverse of the calibration line for the
linear model
|
||
Tests for the intercept and slope of calibration curve -- If both conditions hold, no calibration is needed. |
Before correcting for the calibration line by the equation above, the
intercept and slope should be tested for |
||
Table look-up for t-factor |
The factor, |
||
Quadratic calibration curve |
The inverse of the calibration curve for the
quadratic model
|
||
Power curve |
The inverse of the calibration curve for the
power model
|
||
Non-linear and other calibration curves | For more complicated models, the inverse for the calibration curve is obtained by interpolation from a graph of the function or from predicted values of the function. |